Malvern Instruments
ZEN3600
5년
주장비
분석
화합물 전처리·분석장비 > 입자분석장비 > 입도분석기
2017.10.26.
₩59,400,000
기관의뢰 직접사용
고정형
건별
₩60,000
본 제품은 브라운운동의 원리를 기준으로 한 나노입도 및 제타전위 측정을 위한 장비이다.
본 제품은 에멀전이나 서스펜션 상태의 분산된 나노입도분석이 가능한 장비이다.
또한 에멀전 상태에서의 제타전위측정을 함으로써 분산된 분산안정도 측정이 가능한 장비이다.
Ⅰ. 용도(End-user’s Use)
1. This instrument will be used for normal experiment, training and research purpose.
2. This equipment is an equipment for measuring the particle size of the nano-material.
3. This equipment can measure the zeta potential.
4. This equipment can measure the molecular weight.
Ⅱ. 장비의 구성(Configurations of Goods)
<주장비>
1. Nano Particle Size Analyzer 1 set
2. Standard Accessories 1 set
3. Optional Accessories 1 set
<부속장비>
4. Data Processing Unit 1 set
Ⅲ. 성능 및 규격(Performance and Specification)
<주장비>
1. Nano Particle Size Analyzer
1-1. Feature
1) Simplified sample preparation with little or no dilution for size mearsurement (<40wt%)
2) Comprehensinve range of low volume cells (12ul - 1ml)
3) Non-invasive back scatter technology built-in for increaserd sensitivity and accuracy
of nano particle < 1nm in size and <1000Da in Molecular weight
1-2. Measurement Method
1) Backscatter (more than 173 degree) preferred for concentrated dispersions.
2) Mixed mode measurement(FFR + SFR) method and Electrophoretic mobility measurement
for the Zeta-potential measurement.
3) Organic solvent sample can be measured by Smoluchowski formula and Huckel formula in
analysis software
4) When we measure a particle size, we can make sure Aggregation and dispersion which be
known to quality of particle through measurement quality report and correlogram.
5) Premium system for the ultimate in colloid and protein characterization and also
Microrheological characterization available (optional)
1-3. Optic
1) Laser : 4mW 633nm He-Ne gas laser
2) Detector : Avalanche Photodiode Detector
3) Size measurement : 173 degree back scattering detector
4) Zeta measurement : Laser Doppler method 17 degree
5) Debye Plot for MW measurement
6) Automatic attenuation selection
7) Attenuation level : 1-11 level
1-4. Measurement range
1) Size (diameter) : 0.3 – 10,000nm
2) Zeta-potential : +/- 500mV
3) Size range for Zeta-potentials : 3.8nm to 100 microns
4) Concentration range for size measurement : 0.1ppm, 0.1mg/ml to 40%w/v
5) Conductivity range : 0 to 200mS/cm
6) Minimum Sample volume for Size measurement : 12ul
7) Sample volume for Zeta potential measurement : 20ul with Diffuse barrier method
8) Molecular weight range (Est) : <1000Da to 2x107 Da
9) Temperature range : 0 - 90 degree
1-5. Various cell for measurement
1) Required sample volume for measuring is less than 0.75ml
2) Zeta potential and size can be measured in the same cell
3) Maintenance-free zeta potential cell
4) Universal Dip cell kit for organic solvent based Zeta potential
1-6. Measurement position in sample
1) Variable. The optics are automatically focused to the optimal position in the cell to maximize
the ratio of signal to noise and minimize multiple scattering effects for accuracy.
1-7. Alignment
1) No alignment, Calibration or Maintenance required
1-8. Software
1) Customer report generator
2) Temperature trend analysis
3) Time and measurement trend analysis
2. Standard Accessories
1) Folded capiliary Zeta cell 1 pk
2) 12mm o.d. square polystyrene cuvettes 1 pk
3) 12mm square glass cuvette 1 ea
4) Latex standard (Zeta Potential) 1 ea
3. Optional Accessories
1) Folded capiliary Zeta cell 3 pk
2) 12mm o.d. square polystyrene cuvettes 3 pk
3) Nano Particle Standard Sample, 60nm, 15mL 1 bt
<부속장비>
4. Data Processing Unit
1) Desktop PC
2) 22” LED Monitor
3) Laser Printer