GE Health care
AKTA pure (M)
10년
주장비
분석
화합물 전처리·분석장비 > 분리정제장비 > 달리 분류되지 않는 분리정제장비
2018.03.14.
₩50,583,480
기관의뢰
이동형
시간별
₩45,000
단백질 및 효소등의 분석기능과 주로 biomolecule을 분리하는 전용 system으로서 단백질 분리시간을 단축하기 위한 다양한 벨브시스템과 시료의 확산을 방지하는 detector, two-pump를 사용하는 gradient system을 채용하여 정확한 분석이 가능함. 다양한 단백질과 핵산, 펩타이드 등의 생물 고분자 및 저분자물질을 빠르고 재현성 있게 분석 및 분리가 가능한 장비임
1) High performance dual piston Pump
① Type : Piston pump , metering type
② Operating pressure : 2900 psi (20MPa,200bar)
③ Flow rates range
isocratic mode : 0.001 - 25ml/min (normal range)
50ml (column packing flow )
gradient mode : 0.001 - 25ml/min
④ Increment : 0.001 ml/min
⑤ Pressure limits : programmable upper and lower limit
⑥ Internal volume : < 600ul
⑦ Viscosity : maximum 10cP (5 cP above 12.5 ml/ min)
⑧ Pump head : Titanium alloy
⑨ Flow rates accuracy : +/-1.2 % (conditions: 0.25 to 25ml/min , <3Mpa, 0.8 to 2cP)
⑩ Flow rates reprod. : rsd <0.2%
3) Mixer
① Mixing principle : Chamber with a magnetic stirrer
② Mixer volume : 0.6 1.4 ( included with system) or 5 ml
③ Gradient flow rate range : 0.1 to 25 ml/min
④ Gradient composition : +/- 0.6 % ( conditions : 5 to 95% B 0.6 to )
⑤ Accuracy : 25 ml/min , 0.2 to 2Mpa 0.8 to 2 cP
4) UV Monitor
① Number of monitors : Up to 2
Wavelength range : 280nm
Absorbance range : -6 to + 6 AU
Linearity : ± 5% within 0–2 AU
Operating pressure : 0 to 2 Mpa ( 290 psi)
5) Conductivity Monitor
① Conductivity reading range : 0.01 mS /cm to 999.99 mS/cm
② Accuracy : +/- 0.01 mS/cm or +/- 2% whichever is greater ( within 0.3 to 300 mS/cm )
③ Operating pressure : 0 to 5 Mpa ( 725 psi)
④ Flow cell volumn : 22 ul
⑤ Temperature monitor range : 0 ℃ to 99 ℃
⑥ Temperature monitor accuracy : +/- 1.5 ℃ within 4℃ to 45 ℃
6) pH Monitor
① pH reading Range : 0 to 14
② Accuracy : +/- 0.1 pH unit within pH 2 to 12
③ Operating pressure : 0 to 0.5 Mpa (72 psi)
④ Flow cell volume : 76 ul
⑤ Included flow restrictor ( prevention of air bubble trouble )
7) Pressure Monitor
① Number of sensors : Up to 4
② Placement of sensors
- Standard : The System pressure monitor is located after the System pump
- Included : The Pre-column pressure monitor and the Post-column pressure monitor
are integrated in Column valve
③ Range : 0 to 20 MPa (2900 psi)
④ Accuracy : ± 0.02 MPa or ± 2% whichever is greater
8) Valves
① Type : Rotary valves
② Number of valves : up to 12
③ Functions : standard : injection valve