Varian
2520D
주장비
분석
임상의료장비 > 임상진단/영상장비 > 임상진단용엑스선장비
2013-01-25
60,000,000원
원
The PaxScan 2520D/CL is a third-generation
X-ray imager designed for high-speed dental
cone-beam CT imaging. Based upon the new
Gigabit Ethernet interface standard, or customer-
specified CameraLink standard, images
are displayed instantaneously on a user-supplied
workstation. The housing is lightweight
aluminum.
Receptor Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Amorphous Silicon
Conversion Screen . . . . . . . . . . . . . . . . . CsI, DRZ Plus, or Gd2O2S: Tb
(Kodak Lanex Screen)
Pixel Area
Total . . . . . . . . . . . . . . . . . . . . 19.5 (h) x 24.4 (v) cm (7.68 x 9.6 inch)
Active (Gd2O2S) . . . . . . . . . . 19.3 (h) x 24.2 (v) cm (7.60 x 9.49 inch)
(Csl, Direct Deposit) . . . . . . 19.0 (h) x 23.8 (v) cm (7.48 x 9.37 inch)
Pixel Matrix Total . . . . . . . . . . . . . . . . . . . . . . . 1,536 (h) x 1,920 (v)
Active (Gd2O2S) . . . . . . . . . . . . . . . . . . . . . . . . . 1,516 (h) x 1,900 (v)
(Csl, Direct Deposit) . . . . . . . . . . . . . . . . . . . . . . . . . 1,496 (h) x 1,874 (v)
Pixel Pitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 μm2
Limiting Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.94 lp/mm
MTF, X-Ray . . . . . . . . >48% 1 lp/mm, Gd2O2S:Tb screen (80 kVp)
DQE (0), RQA5 Direct Deposit CSI. . . . . . . . . . . . . . . . . . 70% ±5%
DQE (0), RQA5 Other Scintillator Options . . . . . . . . . . . . . . . ?60%
Energy Range Standard . . . . . . . . . . . . . . . . . . . . . . . . . 40 - 150 kVp
Fill Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57%
Image Capture . . . . . . . . Intel® Pro/1000GT Desktop Adaptor (PCI)
(Customer Supplied)
Epix® CameraLink Adaptor (PCI)
(Optional)
Scan Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Progressive
Data Output/System Interface . . . . . Gigabit Ethernet or CameraLink
A/D Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-bits
Frame Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 10 fps (1 x 1)
15/30 fps (2x2)
(Workstation dependent)
Software
The software release includes ViVA™, a basic application for image
acquisition and viewing on an end-user workstation or laptop runni
Installed on the Rixan Staubli Robot Arm, X-ray Flat Panel detector will be synchronized to the patient’s movements using real-time motion capture.
An image detector, usually composed of selenium or silicon, used in digital radiography to convert x-rays into light or an electronic signal