광전자정밀
OPI-1000
10년
주장비
분석
화합물 전처리·분석장비 > 분광분석장비 > 자외/가시광/적외선분광광도계
2013-03-22
225,000,000원
기관의뢰
고정형
건별
77,000원
본 장비는 농생명 LED 및 모든 광원의 광속, 색온도, 연색성 및 광원의 스펙트럼 등 다양한 광학적 측정이 가능한 장비이다. LED조명의 광학적특성 전광선속, 색 관련(Parameter 등)을 “고효율에너지 인증“ 또는 “KS규격“에 따라 정확하게 측정하여 제품인증을 하는데 사용하기 위함이다.
- 적분구 재질이 알루미늄이어서 강화 플라스틱에 비해 광학적 특성이 우수
- 색보정을 통한 정확한 측정
- 측정값의 불확도 3%미만
1. Integrating Sphere
1) Diameter : 80 inch / approx. 2000mm
2) Sphere material : Aluminium
3) Number of Port : 11ea
Auxiliary lamp, spectroradiometer, DUT power, Temperature sensor,
Photometer 7ea
4) Coating Material : Barium sulfate (BaSO4)
Reflectance ratio : > 95 % for Visible light
5) Light proofing mechanism
3 step light proofing mechanisms
6) Support Mechanism
Aluminum profile and precision guide
7) Uniformity for SRDF(Spatial Response Density Function) is less than 1.5%.
2. DUT Jigs
1) MR type Jig (Gu-5.3 base)
2) Par type Jig (E-26 base)
3) Fluorescent lamp type Jig (G-13 base)
4) Flat panel type Jig (wire)
3. Auxiliary lamp
1) Halogen lamp : 60W
2) Manual switch for connecting to DC power supply
4. Photometer Module System
1) Picoammeter
Current range : 2nA~20mA
Voltage burden : 200uV ( 1mV on 20mA range)
Reading rate : 1000/s
Resolution : 10fA
Digits : 5 1/2
Storage buffer : 2500 points
Interface : RS232
2) Switching System
Channels : 10
Connector type : BNC
Maximum signal level : 30V, 100mA
Contact life : > 106
Contact Resistance : < 1Ω
Contact Potential : < 200μV
Offset Current : < 1 pA
3dB bandwidth : 1MHz
Actuation Time : < 1msec
Channel Isolation : > 1014Ω
3) Class L Detector
Type : Silicon photodiode
Effective area : 100mm2
V(λ) Mismatching factor : f1’ < 1.5% Class L
Thermo-stabilization control : 30℃
4) General Detector : 6EA
Type : Silicon photodiode
Effective area : 100mm2
Spectral range : 190 ~ 1000nm
Dark current : 6 pA
Rise time : 7 μsec
5. Spectroradiometer System
1) Spectrometer
Detector : TE Cooled 1024×122 CCD(back thinned)
Spectral range : 360-830nm
Resolution : 1.0/0.5[nm/ch]
Integration time : 5msec-60sec
Cooling : -10℃
ND Filter : OD 1 to 2
Slit:50 μm fixed
Linearity : ±0.8%
Wavelength accuracy : <±0.3nm
Dominant Wavelength accuracy : ±0.5nm
Luminous intensity accuracy : ±4%
Luminous flux accuracy : ±4%
Data Point Interval : ±0.5nm
Chromaticity (x,y) : ±0.002nm
Stray light Broadband : 1x10^-4
Stray light LED/laser : 1x10^-5
Optical Input : 200μm bundle, 2m long (Original Connection)
Speed : 200scans/sec(at 5msec)
Dynamic range(single scan) : 1500000:1
Spectral Sample interval : 1.0/0.5[nm/ch]
Mechanical Shutter : Yes
AD Converter : 16bit
A/D Rate : 500k(Theory value:1M)
PC Interface : USB2.0(High speed)
2) Optical Fiber
Core Size : 400μm Fused Silica Core
Transmission Range : 250~1100nm
Total Length : 2 meters
Jacket : Metal
Connector : SMA905 Terminated
3) Optical Head
Cosine corrector : Diffuse plate
Connector : SMA905 Terminated for optical fiber
Housing : Anodized aluminium
6. Power Analyzer
1) Input Specifications
Number of channels : 3
Input type : Floating input, Resistance voltage divider, Shunt input system
Voltage ranges: 15/30/60/150/300/600V
Direct current input: 0.5/1/2/5/10/20A
Voltage input resistance : Approximately 2MΩ, 13pF
Direct current resistance : Approximately 500mΩ + 0.1μH
Maximum instantaneous allowed input voltage : 2.8kV peak, 2.0kV rms
Maximum instantaneous allowed input current : 450A peak, 300A rms
Maximum continuous allowed input voltage : 1.5kV peak, 1.0kV rms
Maximum continuous allowed input current : 100A peak, 30A rms
A/D Converter : Simultaneous conversion of voltage and current inputs
16bits resolution
Range switching : Ranges can be set manually, automatically,
or through online controls
2) Measurement Functions
System : Digital sampling(Sum of averages method)
Frequency range : DC, and 0.5 Hz to 100kHz
Crest factor : 3 with rated input
Voltage/Current Accuracy (three months after calibration)
DC : ±(0.2% of rdg + 0.2% of rng)
0.5Hz < f < 45Hz ±(0.1% of rdg + 0.2% of rng)
45Hz < f < 66Hz ±(0.1% of rdg + 0.1% of rng)
66Hz < f < 1kHz ±(0.1% of rdg + 0.2% of rng)
1kHz < f < 10kHz ±((0.07xf)% of rdg + 0.3% of rng)
10kHz < f < 100kHz ±(0.5% of rdg + 0.5% of rng)
Active Power Accuracy (three months after calibration)
DC : ±(0.3% of rdg + 0.2% of rng)
0.5Hz < f < 45Hz ±(0.3% of rdg + 0.2% of rng)
45Hz < f < 66Hz ±(0.1% of rdg + 0.1% of rng)
66Hz < f < 1kHz ±(0.2% of rdg + 0.2% of rng)
1kHz < f < 10kHz ±(0.1% of rdg + 0.3% of rng)
10kHz < f < 100kHz ±(0.5% of rdg + 0.5% of rng)
Line filter function : A low-pass filter can be inserted in the input circuit for
measurement. The cutoff frequency is 500Hz
Temperature coefficient : ±0.03% of range/℃ at 5~18℃ and 28~40℃
Display updating intervals : 0.1/0.25/0.5/1/2/5 seconds
High speed data update : as fast as 10 readings per second
Frequency measurement accuracy : ±(0.06% of rdg)
User calibration capability
Display unit : 7-segment LED
Interface : RS232
7. AC Power supply
Output phase : 1
1) AC output rating
Power/Phase : 1000VA
Voltage Range: 150V/300V/Auto
Voltage Accuracy : 0.2% + 0.2%F.S.
Voltage Resolution: 0.1V
Voltage Distortion: 0.3% @ 50/60 Hz, 1% 15-1KHz
Voltage Line regulation : 0.1%
Voltage Load regulation : 0.2%
Max. Current/Phase for R.M.S : 8A/4A (150V/300V)
Max. Current/Phase for Peak : 48A/24A (150V/300V)
Frequency range : DC, 15 ~ 1kHz
Frequency accuracy : 0.15%
2) Input rating
Voltage range : 90~250V, 1phase
Frequency range : 47~63Hz
Current : 18A max. at 90V
Power factor : Min. 0.97
3) Measurement
Voltage Range.Phase : 150V/300V
Voltage Accuracy : 0.2% + 0.2%F.S.
Voltage Resolution : 0.1V
Current range for peak : 48A
Current accuracy for rms : 0.2% + 0.1%F.S.
Current accuracy for peak : 0.2% + 0.2%F.S.
Power Accuracy : 0.4%F.S.(CF<6)
Power Resolution : 0.1W
Interface : RS232
Operating Temperature : 0 ~ 40℃
Storage Temperature : -40 ~ +85℃
Safety and EMC : CE
8. DC Power supply
DC Output ratings : 0~100V, 0~15A, 1500W
Output Ripple and Noise : 100mV p-p, 12mV rms
Load Effect (change from 10% to 90%) : 17mV, 7mA
Source Effect (change from 85~132VAC input or 170~265VAC input)
: 17mV, 3mA
Programming accuracy : 0.05%+75mV, 0.1%+10mA
Measurement accuracy : 0.1%+150mV, 0.1%+30mA
Load Transient Recovery Time : < 2ms
Output Response Time ( settle to within ±1.0% of the rated output, with a
resistive load) : 0.15 sec for Up, full load, 2.0 sec for Down, no load
Command Response Time : 55ms
Over-voltage protection range : 5~165V with 1.5V accuracy
Front Panel Display Accuracy : 0.75V, 0.05A
Interface : GPIB and USB and Ethernet/LAN
9. Automatic Voltage Regulator
Power rating : 7.5KW
AC input voltage : 1 phase 110, 115, 120, 200, 208, 220, 230, 240
(User Selection)
AC input range : ±15%, ±20%
AC input frequency : 50, 60 Hz (Selection Option)
AC output voltage : 1 phase 110, 115, 120, 200, 208, 220, 230, 240
(User Selection)
AC output range : Typical ±2%
AC outout Feature : Digital Tab Change
AC output frequency : 50 or 60 Hz (Selection Option)
AC output efficiency : >90%
AC output total harmonic distortion : <3%
10. Software
Enable to control all instruments
Enable to set the multi test conditions for optical properties
Enable to display the temperature of inner sphere.
Enable to save data to excel and text format
Enable to save graph to bmp format
Enable to change the measurement units from mW to lm and mol/m2sec
Calculation items : Peak wavelength, center wavelength, centroid wavelength,
half width, 1931/1960/1976 color coordinates, Excitation purity,
correlated color temperature, color rendering index R1~15,
photosythetic photon flux density (PPFD : mol/m2sec).
Windows programming
11. Computer & Rack case
1) Rack case for instruments and drivers
2) Control PC
CPU : Intel Core2 Q9400 , RAM : 2GB
HDD : 1TB, Monitor : 19inch LCD
OS : MS Windows XP
ETC : MS Excel