EM TEST
VDS200Q
10년
옵션/액세서리(보조장치) (주장비:외부 전자파 환경측정시스템)
시험
전기·전자장비 > 측정시험장비 > 전압/전류/전력측정시험장비
2022-09-22
353,825,280원
기관의뢰
고정형
일별
[Hr] 225,000원
전기적 환경 시험기 (Electrical Environmental Test System)
- Battery Simulator
- Wave Generator
- Power Fail Simulator
- Audio amplifier with the internal signal generator
- Ford EMC-CS-2009.1 Test Box
- Remote-controlled DC Source
- Coupling Network for DC-10615
- Coupling network for Ford EMC-CS-2009.1 CI 250
- System Accessories
A. Transient Immunity Test System
- Test standard : ISO 7637-2/-3 and ISO 16750-2
1. Simulator 1 : 1 set
1.1. Built-in coupling networks
- DUT voltage : 70 V(Max) or more - DUT current : 50 A or more
- Peak Current : 100 A for 500 ms or more
1.2. Electrical Fast Transient (for Pulse 3a/3b)
- Test Voltage : 30 V ~ 1 000 V ± 10 % or more - Rise time : 5 ns ± 1.5 ns
- Pulse duration : 150 ns ± 45 ns or 100 ns ± 10%
1.3. Pulse 1
- Rise time : 1 us, 3 us or more
- Pulse duration : 1 ms, 2 ms or more - Internal resistor : 10, 50 ohm or more
1.4. Pulse 2
- Rise time : 1 μs or more
- Pulse duration : 50 us, 300 us or more - Internal resistor : 2 ohm or more
1.5 Should generate JASO D-001 pulse A2, B2, D2 and NiSSAN 28400 NDS 03, Pulse B2,
NISSAN 28400 NDS 07, pulse C8, C50, C300
2. Simulator 2 : 1 set
2.1. Freestyle mode
- Test Voltage (O.C) : 20 V ~ 200 V ± 10 % or more - Rise time : 0.09 ~ 10 ms or more
- Pulse duration : 30 ms ~ 1 200 ms or more - Internal resistor : 0.5 ohm ~ 10 ohm or more
2.2. Clipped load dump
- Clipped voltage : 15 V ~ 99.5 V or more
3. Simulator 3 : 1 set
- Output range : -20 V ~ + 70 V or wider
- Output current : 0 A ~ 50 A, Continuous or more - Peak current : 150 A for 200 ms or more
- Frequency range : DC to 150 kHz or wider
- Supply voltage : 3-phase 200/400 V ± 10%, L1, L2, L3, PE - Source impedance : < 10 mohm or more
- Should generate ISO 7637-2 Pulse 2b and Pulse 4, ISO 16750-2
4. System Software : 1 copy
- Should control all the functions of the system
- Should have libraries for international and manufacturer standards
B. Transient Emission Test System
- Test standard : ISO 7637-2
1. Electric specification: 1 set
- Battery current: 100 A or more
- Inrush current: 300 A for 200 ms or more - Battery voltage: 0 V ∼ 60 V(DC)
1.1 Electric switch: 1 set
- Switching time: 300 ns ± 20 %
- Voltage drop: below 1 V at 25 A , 2.1 V(typ.) at 100 A - Transient voltage protection: includes
1.2 Mechanical switch: 1 set
- Voltage rating: 400 V or more
1.3 Artificial network
- Should comply impedance curve as per ISO 7637-2
C. Radiated Immunity Test System
1. Line impedance stabilization network: 2 ea
- Frequency range: 100 kHz ∼ 150 MHz or wider - Continuous current: 200 A or higher
- DC voltage: 1 000 V or higher
- Impedance: (5μH + 1 Ω)∥50 Ω within ±10 %
2. High Gain Horn Antenna: 1 ea
- Frequency range: 4 GHz ∼ 8 GHz or wider - Power Handing: 500 W or more
- 3 dB Beamwidth: Datasheet Necessary
3. High Power RF Cables: 1 lot
: dc to 6 GHz : min 3 kW
-Frequency range -Power rating
D. Test Accessories
1. Calibration Set for Load Resistor : 1 set
-Built-in resistor values -Resistor tolerance
2. Capacitive Coupling Clamp: 1 ea -Max, voltage
-DUT cable
3. Capacitor Box : 1 ea -Capacitor Value
: 0.5, 1, 2, 4, 10, 20, 50 ohm, selectable : ± 1 %
: 1 kV
: up to 40 mm diameter
: Fast transient pulse 100 pF, Slow transient pulse 0.1 uF