Waters
ARC HPLC system
10년
주장비
분석
화합물 전처리·분석장비 > 분리분석장비 > 액체크로마토그래피
2022-08-31
64,061,730원
기관의뢰 직접사용
고정형
시간별
[Hr] 10,000원
1. Enhance performance without redeveloping your method
2. Run methods with greater robustness over traditional HPLCs
3. Easily replicate, adjust, and enhance existing HPLC methods with low analyte carryover, high injection precision, and high back pressure tolerance
4. Seamlessly receive transferred methods from other HPLC platforms, preserving analyte retention time and reproducibility of your analysis
5. Support compliance with Empower Chromatography Data System (CDS), the gold standard, compliant-ready software that facilitates data review, eliminates burdensome compliance documentation, and improves the overall quality of laboratory data
1. High Performance Liquid Chromatography
1) Solvent Manager
(1) Type : Quaternary Precise
(2) Number of Solvents : One to four
(3) Integrated vacuum degassing : 4 chambers
(4) Operating Flow rate Range : 0.001 – 5.00 mL/min
(5) Gradient Profiles : Eleven(11) gradient curves
(6) Maximum Operating Pressure : 9,500psi
(7) Flow Precision : ≤ 0.075 % RSD
(8) Flow Accuracy : ≤ ±1 %
(9) Composition precision : ≤ 0.15 %
(10) Delay volume : ≤ 1000uL
2) Sample Manager
(1) Number of Sample vials : 96 vials(2ml vial holder)
(2) Sample Carryover : < 0.002%
(3) Sample delivery precision : <0.5 % RSD
(4) Injection volume range : 0.1 – 50uL
(5) Sample Temperature Control : 4.0 to 40.0°C
3) Column Heater
(1) Column Temperature Control : 15.0 °C below ambient to 65.0 °C, 0.1 °C increments or wider
4) Solvent Tray Module : Tray accommodates up to four(4) one-liter eluent bottles
5) System Start Up Kit : System Test Mix and Column
2. UV Detector
1) Wavelength Range : 190~700 nm
2) Wavelength Accuracy : ≤ ±1 nm
3) Bandwidth : ≤ 5nm
4) Flow cell volume : ≤ 16.3uL
5) Baseline Noise : ≤ 5×10-6 AU
6) Drift : ≤ 1×10-4 AU
7) Light source : Deuterium lamp (Warranty 2000 hour)
3. FLD Detector
1) Wavelength Range : 200 – 890 nm
2) Wavelength Accuracy : ≤ ±3 nm
3) Bandwidth : ≤ 20 nm
4) Flow cell volume : ≤ 13uL
5) Baseline Noise : ≤ ±10 X 10-6 AU or Raman (H2O) > 3000 (noise reference measured at dark value)
6) Light source : Xenon Flash lamp (Warranty 2000 hour)
4. Chromatography software or equivalent
1) Full 64 bit OS System
2) The software include data acquisition rate, instrument control, data analysis
3) An auto-integration function
4) Quantification is based on %, Normalized %, External standard, External standard %, Internal standard and internal standard % calculations calculated on either peak area
5) System Suitability tests the following:
- Method validation
- System performance
- Reproducibility
- Tracking and plotting trends
- Processing and reporting
6) Include Audit Trail
5. Data System
1) CPU : Intel® Core™ i7-9700 Processor (3.0GHz 12MB Cache8Cores 8Threads)
2) RAM : 16GB DDR4-2400 DIMM 1TB Serial ATA Hard Drive
3) HDD : 1TB 7.2K rpm SATA 6.0Gb/s 3.5“ Hard Disk Drive 1G Ethernet Networking LAN
4) O/S : Windows 10 Pro / 64bit (English)
5) Super Multi DVD
6) Monitor : 24“ TFT
7) Printer : Laserjet(black and white)
6. Accessories
1) UV IQ/OQ test solution 1 ea
2) UV Lamp 2 ea
3) Xenon Lamp 2 ea
4) outlet valve 3 ea
5) purge valve 3 ea