Infors HT
Multitron
10년
주장비
분석
화합물 전처리·분석장비 > 바이오제조/분석장비 > 배양장치
2020-04-02
56,289,110원
원
해당장비는 일정 온도에서 일정 회전속도로 진탕시켜 동·식물 세포의 진탕배양, 호열성균 배양, 각종 효소촉매의 반응에 사용하는 배양기의 일종으로서 바이러스벡터를 생산하기 위해 부유세포의 배양에 사용되는 장비다.
1. Shaker
1) Speed range(Orbital) : 30 rpm to 400 rpm or wider
2) Accuracy : ±0.1 rpm or less
3) Motion type : Magnetic type or Belt drive type
4) Diameter/Stroke : 25 mm or wider
5) Loading : Max. 25kg or more
2. Temperature
1) Ambient -10°C to 65°C or wider (with humidity control)
2) Accuracy absolute: ±0.30°C or less (37°C)
3) Control : PID, digital.
4) Probe : RTD, PT100.
5) Heater : Two finned stainless steel heater 1000W or more
6) Refrigerator : 155 to 270W or wider
7) Air circulation : 300 m³/h or wider
3. Humidity
1) Principal : Electrically heated evaporation basin.
2) Range : 85% rH or more
3) Accuracy : ±2% r.h or less
4) Control : PID, digital.
5) Heater : Waterbath (300 Watts or more), Window (100 Watts or more)
6) Door heater : The doors are heated to avoid condensation on glass.
4. Integrated CO2 Control
1) Measuring Range : 0 ~ 20 % or more
2) Accuracy : ±0.02% Relative or less
3) Sensor : Infrared, NDIR.
5. Lighting : LED or Halogen
6. Sterilization
1) Chemical : As the complete inner body is made out of stainless steel, The incubator can be easily sterilized with chemical.
7. Safety
1) Automatic restart : After a power failure, the machine restart automatically when power is restored.
8. Self monitoring system : All parameters are checked and switched off in case of fault.
9. Alarm network : An alarm network interface can be installed to connect the alarms to an external monitoring switchboard.
10. Door switch : Shaker unit, fan and heater stop automatically by door opening.
11. Maintenance : With cooling (Cleaning dust from the condenser).
12. Chamber Volume (empty) : 350L or more
13. Culture capacity : 100ml Flasks (70ea or more), 150ml Flasks (50ea or more),
250ml Flasks (40ea or more), 2L Flasks (9ea or more)
14. Power supply
1) Mains : 220-240 VAC, 50-60 HZ.
2) Consumption : 1300W without cooling, 2000W with cooling.
3) Overvoltage protection : The machine is protected against voltage fluctuation in the supply.