JAI Co., Ltd.
LC-20AP Prominence Series
10년
주장비
분석
화합물 전처리·분석장비 > 분리분석장비 > 분취용크로마토그래피
2019-12-18
90,023,780원
기관의뢰
고정형
일별
30,000원
◦ 천연물 원재료를 이용한 주요 유효성분의 분석 및 천연물의 다양한 미지의 유효성분 분석에 필수적인 장비임
◦ Recycling method를 이용하여 시료 주입을 재순환함으로서, 천연물 등에 포함된 미량의 유용성분을 고순도 대용량으로 분리정제 가능
◦ 바이오테라피용 식의약 및 화장품 소재의 물질탐색과 분리분석 및 개발제품의 품질관리를 위한 유효/유용 및 기능성 성분의 분리정제와 정량분석에 필수적
◦ PrepLC System
1) Solvent Delivery Unit
(1) Pump type : Reciprocating double plunger type or Parallel-type double plunger
(2) Quaternary gradient & Isocratic system
(3) Max. Pressure : ≥ 20Mpa
(4) Flow rate range : 0.01mL/min ~ 30.0mL/min or more
(5) Flow rate Precision : ≤ 0.2% RSD
(6) Automatic plunger rinsing
2) Sample Injector
(1) Manual Injector
- Rheodyne manual operation
- Sample loop vol. : 10mL
(2) Repeat Injector
- With non-diffusion function recycling system
- Recycling-start time : adjustable from 1 to 999min or more
- Auto recycler 1 to 99times repetition or more
3) Detector
(1) UV/VIS Detector
- Light source : Deuterium (D2) & Tungsten (W) lamp
- Wavelength range : 200 ~ 840nm
- Bandwith : ≤ 8 nm
- Wavelength accuracy : ≤ ±1 nm
- Wavelength reproducibility : ≤ ±0.5 nm
- Noise : ≤ 5×10-5 AU
- Drift : ≤ 1×10-3 AU/hr
(2) Refractive Index Detector
- Light source : LED or Tungsten lamp
- RI range : 1.00 ~ 1.75 RIU
- Measurement range : 1.5×10-6 to 3,000×10-6 RIU
- Flow rate : ≥ 100mL/min
- Noise : ≤ ±1.5 × 10-8 RIU
- Drift : ≤ 3.2 X 10-7 RIU/hr
- Flow cell volume : ≤ 9uL
- Temp. accuracy : ≤ ±0.01℃
◦ Fraction collector
1) Collection mode : manual, peak mode or more
2) Max. number of fractions : ≥ 33 bottles or 54 test tubes(ø35)
3) Collection method : manual mode, auto mode
4) Maximum flow rate : ≥ 60mL/min
◦ Chromatography Manager & Controller
1) Chromatography Manager
(1) Auto repeat program : Repeat program automatically takes up a sample and repeat the solution into th solvent suction line at preset intervals.
(2) Auto clean up system is automatically solvent cleaning. There is column reject valve. So inner line cleaning is easily and column, repeat line, recycle line and all body are automatically cleaning.
(3) Application flexibility with HPLC system control capability
(4) This manager can be chromatography data storage and print
(5) exported file’s format of txt. pdf. raw. and mat or more
(6) Easy operation with wizard function
2) Controller