Beckman coulter
XE-100
9년
주장비
분석
화합물 전처리·분석장비 > 분리정제장비 > 원심분리기
2019-04-01
82,079,230원
원
핵산 (DNA, RNA) 및 단백질 분리에 사용. 단백질 및 결합 연구 및 바이러스 격리. 분자량 결정 관련 연구 및 Submicron 및 subcellular 분별 연구.
1. Ultracentrifuge System Performance
1) Full microprocessor control system.
2) The system uses a digital display and interactive menu driven operation.
3) Automatic operation-preheat, precool, dry cycle, over-temperature protection.
4) The HEPA filter is available to protect user from biohazards.
5) Offering regulatory compliance including GLP, GMP and EQ; Design Qualification (DQ), Installation Qualification (IQ), Operation Qualification (OQ) and performance Qualification (PQ).
6) Top loading swinging bucket rotor provides improved performance, ease-of-use and safety.
7) Touch-screen display - Displays more data. Easy to read and navigate. Viewable from across the lab.
8) Energy efficient or less - Regenerative braking returns energy to local circuit reducing utility costs.
9) Zonal and continuous flow operation screens - Intuitive, graphical step-by-step interface simplifies workflow
10) Multilingual software - Includes nine languages for error-free operation
11) Improved power tolerance - Handles a wider tolerated range of 180 VAC to 264 VAC.
12) Quiet-drive technology - Operates at less than 51 dBA. or less
13) Context-sensitive on-screen help - Find answers easily. Reduce training time and user error.
14) This System must be compatible with existing ultracentrifuge rotors
1. Ultracentrifuge System Specification
1) Max. speed : 100,000 rpm or more
2) Max. force : 802,000 g or more
3) Speed control : ±2 rpm or less
4) Set temperature : 0 to 40℃
5) Temperature control/display : ±0.5℃ of set temperature or less
6) Vacuum : 5 micron (0.665pa) or less
7) Vacuum display : Digital 1 micron or less
8) Vacuum system : Moisture purging vacuum system
9) Heat output : 1.0 KW (3,400 Btu/h) or less
10) Sound level (3ft from instrument) : 51 dBa or less
11) Timer : Up to 99 hrs. 59 min. holds or more
12) Drive warranty : 10 years or more
13) Drive cooling : Air cooled, no compressor for vacuum-encased induction direct drive system
14) Refrigeration system : Thermoelectric-no CFCs, ODCs
15) User-settable programs : 9 user programs with delayed start capability or more
16) Programmed acceleration/deceleration rates : 10 Acc./11 Dec or more
17) Single tube sample volume range : 0.5ml to 250ml
18) Energy efficient : Uses the same amount of power as a 60-watt light bulb while idling.
19) Multilingual Software - Includes nine languages or more for error-free operation. (Included Korean)
20) Zonal and continuous flow operation
2. Fixed Angle Rotor
Max. speed : 70,000 rpm or more
Max. force : 504,000 g or more
Capacity : 8 x 39 ml or more
Tube Vol. Range w/g-Max (ml) : 4.0~39.0mL or more
3. Ultra Swinging Bucket Rotor
Max. speed : 28,000 rpm or more
Max. force : 141,000 g or more
Capacity : 6 x 38.5 ml or more
Tube Vol. Range w/g-Max (ml) : 8.4~38.5mL or more