SHIMADZU
HS-20
10년
주장비
시험
화합물 전처리·분석장비 > 분리분석장비 > 달리 분류되지 않는 분리분석장비
2017-11-24
50,094,000원
기관의뢰
고정형
건별
18,000원
A. 용도 (Features)
1. For education, quality control and research work.
2. Analysis of wide fields of organic and inorganic compounds such as gases, liquids and solids.
3. The installation and operation should be carrier out by the bidder/supplier’s responsibility at
end user’s site.
B. 구성 (System Configuration)
1. Headspace Auto-sampler (주장비 1set
2. Accessories
C. 사양 (Specification)
1. Headspace Auto-sampler
1) Sample Injection Method : Sulnert sample loop 1mL (standard) ; 0.2mL, 3mL (optional) or
trap
2) Number of Vials : 90
3) Vials : Outer dia. 22.5mm × 79mm (20mL) ; outer dia. 22.5mm × 46mm (10mL) ; can be
combined
4) Leak Check : All-vial automatic check
5) Optional Barcode Reader : Optional, can read 6 types of barcodes
6) Vial Warming Temperature : Room temperature +10 to 300°C (Settings are 0 to 300 °C, in
1°C units, with an accuracy of ±0.1°C)
7) Sample Line Temperature : Room temperature +10 to 220°C or 150 to 300°C (set in 1°C
units, accuracy of ±0.5 °C)
8) Transfer Line Temperature : Room temperature +10 to 350°C (set in 1°C units, accuracy of
±0.5°C)
9) Trap : Inner dia. 2mm × 100mm, Sulfinert tube, Filler TenaxTA (standard),
Carbopack + Carboxene (optional)
10) Trap Cooling Temperature : -30 to 80 °C (set in 1 °C units, accuracy of ±1 °C)
For a sample line at 250 to 300 °C, room temperature - 30 °C
For 150 to 250 °C, room temperature - 40 °C
For 150 °C or less, room temperature - 50 °C
11) Trap Heating Temperature : 0 to 350 °C (set in 1 °C units, accuracy of ±1 °C)
12) Carrier Gas Control : Electronic control via AFC built into GC
13) Vial Pressurized-Gas Control : Electronic control via APC built into GC
14) Carrier Gas and Vial Pressurizing Gas : High-purity helium or nitrogen
15) PC Interface : USB
16) Control Software : Operates collectively with LabSolutions LC/GC (FDA CFR 21 Part 11
compliant)
17) Software Operating Environment : Windows XP, Windows Vista, Windows 7 (32/64-bit)
18) Guaranteed Operating Environment : 15 to 30°C, humidity up to 70 % RH (performance
guaranteed at 18 to 28 °C with temperature fluctuations within ±1.3 °C)