Toyo system Co.,Ltd.
TYS-31TU00SC
5년
주장비
분석
전기·전자장비 > 측정시험장비 > 전지/충방전측정시험장비
2017-06-30
48,592,570원
원
○ 본 장비는 일정 전류 및 일정 전압을 유지하며 개발된 소재를 적용한 전지를 충전 혹은 방전하여 다양한 전기화학적 특성 (방전 용량, 충/방전 속도에 따른 출력특성 및 충/방전 횟수에 따른 전지의 수명특성)을 측정하는 장비임.
○ 본 장비를 통해 도출된 분석 결과에 따라 개발된 소재의 추가적인 생산 또는 수정여부가 판단되기 때문에, 특히 상업화를 위한 소재 개발에 있어 그 중요도가 매우 높은 장비임.
1. Control method
- Charge mode : CC, CC/CV
- Discharge mode : CC,CC/CV, CP
2. Output Voltage Range : -2 ~ +5[V]
3. Output Current Range : 0 ~ 100[mA]
4. Output Wattage Range : 0 ~ 500[mW]
5. Output Function
1) CV
- Setting resolution : 1[mV] below
- Stability : Max. +/-5[mV] below
2) CC
- Setting range: 1~100[mA]
- Stability : Max. +/- 0.5[%] below of low current range (1 mA)
6. Measurement
1) Voltage
- Measurement points : 1 point/ch
- Measurement range : -2 ~ +5[V]
- Maximum resolution : 5[V]/0.5[mV], 2[V]/0.2[mV]
- Accuracy : Max. +/-0.1% below of Full Scale
2) Current
- Measurement points : 1 point/ch
- Measurement range : 0 ~ 100[mA]
- range/ resolution :
100[mA]/12[uA], 10[mA]/7.5[uA], 1[mA]/50[nA]
- Stability : Max, +/- 0.5% below of full scale (each range)
7. Protection Function
1) Over / Under Voltage Protection
- When a voltage value is set up in binary code with the Digital switch for a maximum voltage value
setup and a battery voltage value exceeds a maximum voltage value, an output is stopped without a
microcomputer(Farmware)in a hard circuit.
2) Doubleness of a battery voltage measurement system
3) Over Current Protection : When an over-current occurs in a secondary side, an output is stopped in a
hard circuit. Conditions of operation: 130% of a setting value and 100ms or less
4) AC input over-current protecion : AC line is intercepted by the circuit protector.
5) Reset by communication interception : Host PC and this unit are performing control and measurement
always communicating. When communication stops in a host’s abnormalities after [of a there] 1 minute
this unit carries out self-reset and stops Charge/ Discharge.
6) Countermeasure of the abnormal constant current value : If the measured CC value is Lower than 10% of
the set CC value, Control PC will display the alarm message and the examination of CH is forced to terminate.
7) Countermeasure of the abnormal constant power value: If the measured CP value is lower than 10% of
the set CC value, Control PC will display the alarm message and the examination of CH is forced to terminate.
8. Another Function
1) patterns (more than 80 patterns)
50 modes more/1 pattern, 999 cycles more/1 mode, Loop mode
2) Every channel can be operated individually and independently.
3) Graph: Output is the form of graph can be prepared with the Received data as follow,
- X-axis : Time, Capacity, Cycles, Voltage, Capacity/Weight, Energy
- Y-axis : Voltage, Current, Temperature(op), Pressure(op), Capacity, Power, average Voltage, Average
temperature(op), average Pressure(op), peak voltage, Peak temperature(op), Peak pressure(op)
IV. 기타조건(REMARKS)
1. Installation and test-run should be performed by the authorized Engineer at customer site free of charge.
2. Training : On site training for operation,
3. Warranty : 12 months after installation.
- output current range가 100~0.0001 mA로 미세 전류까지 조절이 가능하여 정확한 평가를 진행하는데 적합