LMT
GO-DS 2000
10년
주장비
시험
광학·전자 영상장비 > 광파발생/측정장비 > 달리 분류되지 않는 광파발생/측정장비
2012-06-28
531,298,938원
기관의뢰 직접사용
고정형
시간별
114,100원
1) Rotating Mirror Type Goniometer
2) Luminous Intensity Distribution Measurement
3) Absolute Total Luminous Flux Measurement
4) Type C-Photometry & Type B-photometry
5) Conforming to International Standards
- IESNA LM79-08 - IESNA LM35-02
- IESNA LM31-95 - IESNA LM41-98
- IESNA LM42-72 - IESNA LM46-98
- CIE 43:1979 - CIE 69:1987
- CIE 70:1987 - CIE 84:1989
- CIE 121:1996 - CIE 127:1997
- EN 13032-1, etc
장비 구성 1. Main system 1) Mirror - Size : 1900 mm × 2690 mm × 6 mm - Quality : Minimal distortion of the reflected image - Axis : 0°~180°, 0°~360°, Resolution 0.1°, - Motor : DC motor 2) Luminaire - Axis : 0°~360°, Resolution 0.1° - Angle measurement : Using absolute angle encoder - Motor : DC motor - Maximum test items : 2000 mm in diagonal, 50kg, and 800 mm height. - Controlling height : Motorized 3) Measurement system , - Photometer(Photo cell) : V(λ) match characteristic f1'<1%, Range 0.0001~80,000 ㏓ - Photocurrent preamplifier - Colorimetric System : Range 380~780 nm, Resolution 1 nm, Integration rate 13 ms ~ 100 s, In accordance with IES LM 79-80 4) Lighting Tunnel, - Distance : 24.41 m - Dimension : Width 2 m × Length 2 m × Hight 15 m 5) AC Power supply and Power analyzer - Voltage range : 150 V/300 V - Frequency range : 45~500 ㎐ - Power capacity : 1500 VA - Stability of supply voltage : 0.2 % - Stability of frequency : 0.5 % - Power Analyzer : Yokogawa Model WT 3000, DC Accuracy 0.05 % of reading + 0.05% of range, AC Accuracy 0.01 % of reading + 0.03% of range 6) Computer system and software - Computer system : Intel pentium processor, Window XP, 4GB RAM and 600GB, 21" LCD monitor - Control software : Luminous Intensity Distribution, Luminous Flux, Monitoring the stability of test samples, Standard Deviation, Stray Light subtraction - Photometric data file format : ASCII file format, EULUMDAT formats, IES formats, CIE Publication 102 formats, EN 13032-1 formats