Yokogawa
AQ6375
9년
주장비
계측
전기·전자장비 > 신호발생장비 > 달리 분류되지 않는 신호발생장비
2012-02-24
94,900,000원
기관의뢰
이동형
건별
200,000원
1) 2㎛ spectrum measurement
2) High wavelength accuracy and resolution
3) High sensitivity measurement
4) Lasers and optical devices test/optical fiber communications
5) Passive optical device measurement&N&T&I&S&1) Wavelength range: 1200 to 2400 nm
2) Wavelength accuracy: ±0.50 nm (Full range)
3) Wavelength repeatability: ±0.015 nm (1 min.)
4) Wavelength resolution setting: 0.05, 0.1, 0.2, 0.5, 1 and 2 nm
5) Minimum sampling resolution: 0.02 nm
6) Number of sampling: 101 to 50001, AUTO
7) Level sensitivity: -70 dBm (1800 to 2200 nm),
-67 dBm (1500 to 1800 nm, 2200 to 2400 nm),
-62 dBm (1300 to 1500 nm)
8) Maximum input power: +20 dBm/channel (full wavelength range)
9) Maximum safe input power: +25 dBm (Total input power)
10) Level accuracy: ±1.0 dB
11) Level linearity: ±0.05 dB
12) Polarization dependence: ±0.1 dB
13) Dynamic range: 45 dB (Peak ±0.4 nm, resolution: 0.05 nm),
55 dB (Peak ±0.8 nm, resolution: 0.05 nm)&N&T&I&S&2㎛ (1.2 ~ 2.4㎛) 중적외선 대역의 광수동 소자의 성능 분석을 위한 스펙트럼을 정밀하게 측정. 중적외선 대역의 광능동 소자 및 레이저 등의 입출력, 흡수/방출 특성 스펙트럼을 측정
- 고분해능 광스펙트럼 분석기는 2㎛ 대역의 장파장 영역에서 광 능동/수동소자의 광학적 투과/손실 특성을 측정할 수 있는 스펙트럼 분석 장비임.
- 2㎛ 대역의 고출력 레이저의 스펙트럼0.05 nm 분해능으로 정밀하게 측정할 수 있는 장비임