A. Specification 1. Method : - Laser diffraction using Fraunhofer and Mie scattering theory including back scattering. - The transmitter module includes He-Ne laser combined with blue solid state light. Optics provides improved resolution in the sub-micron region with a new blue solid state light source and additional detectors. 2. Measuring range : - 0.02 to 2000 microns in a single range. - This system does integrate back scattering into its optical system and measures up to 0.02microns. 3. Resolution : Results reported in 100 size classes. 4. Measuring sample : Emulsions suspensions dry powders. 5. Sampling method : a. Powders in liquid suspension b. Emulsion measurement : Emulsion can be used to suspend powders for analysis c. Direct dry power : A single lens optical system is used for the analysis of both wet and dry samples and no lens changing when switching from wet to dry. 6. Single sample dispersant unit Incorporating variable power ultrasonic dispersion and mechanical stirrer and recirculation pump with independently variable speed. Ultrasound power can be changed during a measurement. Provision is made for the use of a wide range of sample dispersion units which automatically recognised and one-handed interchange.Standard Accessory a. Window tool 1 ea b. Lens Tissue 1 ea c. Tygon tubing 1 ea d. Fuses 1 ea B. Accessories 1. Window and seal
Laser Particle Size Analyzer Malvern model Mastersizer 2000 for 220V AC 60Hz #APA5025 with #APA5004