Thermoscientific
noran system7
10년
주장비
분석
전기·전자장비 > 신호발생장비 > 달리 분류되지 않는 신호발생장비
2011-09-09
80,500,000원
기관의뢰
고정형
건별
45,000원
1) Digital Acquisition Electronics
1-1. Intelligent analyzer electronics
1-2. Dedicated 32-bit microprocessor or more
1-3. High speed, bi-directional data transfer link
1-4. True Digital pulse processing electronics using digital
signal processors (DSPs) for improved resolution,
increased x-ray count rate with reduced dead time
1) Peltier Cooled DRY(No-fan) type Detector featuring
SDD(Silicon Drift Diode)
2) Single Window / 10mm2 SDD Crystal
3) Resolution : 132eV @ Mn standard or better.
4) Detectable elements : Be(4) - U(92)
3-1. Spectral Match software for least squares fitting and
matching of spectra to spectral database.
3-2. X-ray phase analysis software
3-3. Feature Sizing and Chemical Typing software. Includes
size, shape and chemical typing analysis of features in
an electron image.
3-4. Beam current software and interface hardware for SEM.
1. EDS Detector type should be Silicon Drift Diode with Peltier
cooling method.
2. EDS system should provide X-ray Interface Kit for KOPTI
SEM 'CX-100S'
3. X-ray microanalysis software should provide' chi-sqd' value
for a Quantitative Data accuracy.
4. EDX Identification software should provide automatic separate
function for Overlapping peaks.
5. Once EDX counts acquired, the complete data set should
stored on EDX mainframe for analysis at any time
6. EDX System should provide Real-time-display of data being
acquired